3Shape TRIOS S1A-10 Instrukcja konfiguracji - Strona 5

Przeglądaj online lub pobierz pdf Instrukcja konfiguracji dla Sprzęt dentystyczny 3Shape TRIOS S1A-10. 3Shape TRIOS S1A-10 13 stron.

Step 1: Move the arm to the top-most position.
Step 2: Insert the tool into the hole on the bottom side of the arm,
located approx. half way along the arm.
Step 3: Turn the tool clockwise to increase friction, anticlockwise
to reduce friction.
Step 4: Adjust friction so that the monitor is steady
(does not sink, does not go up) in the following scenarions:
• Scanner is positioned in scanner holder
• Scanner is not positioned in scanner holder
IMPORTANT NOTICE
Maximum current available for the USB port is 200 mA.
System with pod (S1P-2, S1P-4) furthermore comprises:
Pod with electronics, PS1-2 (Figure 5):
• A mount for the scanner
• Several connection sockets. On one side, there is a single socket
for the scanner. On the other side, there are two connection sockets
for the Ethernet cable to the PC and for the external medical power
supply, respectively.
• A USB license dongle (access to 3Shape software)
4.7 Setting up TRIOS
It is recommended that unpacking, assembly and connection of the TRIOS
IOS system shall be performed by TRIOS authorized technical service
personnel.
CAUTION
If the system has just been delivered from a cold
environment, let it adjust to room temperature to avoid
internal condensation and to allow for proper calibration.
CAUTION
Examine all parts of the system. Look for physical damage,
loose parts or signs of wear that could interfere with proper
use and functionality. Contact your service provider if there
is visible damage or malfunctioning.
System S1A (MOVE/MOVE+ cart) only
Please follow the steps below for assembling TRIOS IOS system (Figure 1&2):
Before you start: Follow the instructions in the packaging to assemble
the foot onto the the MOVE/MOVE+ cart. Tools are provided.
Step 1: Place the MOVE/MOVE+ cart optimally in relation to the patient
chair and your work position.
Step 2: Attach the protection tip to the TRIOS scanner
if not already in place. (S4)
Step 3: Place the scanner on its designated scanner mount
on the MOVE/MOVE+ cart (S13).
Step 4: Connect the scanner cable to the scanner connection socket
on the MOVE/MOVE+ cart behind the screen below the hinge (S16).
• To connect - Push the plug into socket. There is a red orientation
dot on the socket and plug that should match.
• To disconnect - Pull on the outer release sleeve.
Step 5: Connect the MOVE/MOVE+ cart end of the power cable
to the socket at the bottom of the computer unit (S18).
Step 6: Connect the power cable to a grounded power outlet. Ensure
that the MOVE/MOVE+ cart is placed so it does not block the switch on
the mains socket-outlet. To isolate the MOVE/MOVE+ cart and thus the
scanner from supply mains, unplug the flexible power cord from mains.
Step 7: Connect the LAN cable to the MOVE/MOVE+ cart's LAN
connection at the bottom of the computer unit (S19).
System S1P (pod) only
Please follow the steps below for assembling TRIOS IOS system
(Figure 5 & 6):
Before you start: Place the pod on a flat and horizontal surface optimally
in relation to the patient chair and your work position. Attach the protection
tip to the TRIOS scanner if not already in place. It is easiest not to place
the scanner onto the pod at this time yet, because you may find it easier
to lift the pod while connecting cables in the following.
Step 1: Connect the Ethernet cable to the pod. The cable goes into the
left connection socket on the back side of the pod (the side that has two
connection sockets). If you need to release the Ethernet cable, press on
the small tab on the top of the connector.
Step 2: Connect the power supply to the pod. This cable goes into
the connection socket next to the Ethernet cable described in the
previous step.
• To connect - Push the plug into socket. There is a red orien tation
dot on the socket and plug that should match.
• To disconnect - Pull on the outer release sleeve.
The indicator light next to the scanner connection socket will light green.
Step 3: Connect the scanner cable to the scanner connection socket on
the front side of the pod (the side that has only one connection socket).
• To connect - Push the plug into socket. There is a red orien tation dot
on the plug that should point upwards.
• To disconnect - Pull on the outer release sleeve.
Now all connections to the pod are established, and you can place
the scanner onto the pod.
Step 4: Connect the other end of the Ethernet cable to the PC (if no
Ethernet ports are available in the PC use an Ethernet to USB adapter)
Step 5: Connect the power (mains) cord to the power supply. The other
end of the power (mains) cable goes into a grounded power outlet
(not shown in the Figure). To isolate the pod and thus the scanner from
supply mains, unplug the power cord from mains.
5