CONTEC MEDICAL SYSTEMS 2.3.10.00008 Podręcznik użytkownika - Strona 14
Przeglądaj online lub pobierz pdf Podręcznik użytkownika dla Sprzęt medyczny CONTEC MEDICAL SYSTEMS 2.3.10.00008. CONTEC MEDICAL SYSTEMS 2.3.10.00008 14 stron. Pulse oximeter probe
CONTEC MEDICAL SYSTEMS CO.,LTD
Recommended separation distances between portable and mobile
portable and mobile RF communications equipment and the
The pulse oximeter probe is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the pulse oximeter probe can help prevent
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the pulse
oximeter probe as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Address: No.112QinhuangWest Street , Economic & Technical Development Zone, Qinhuangdao, Hebei
Province, 066004, P.R.China
Tel: +86-335-8015430
Fax: +86-335-8015430
E-mail: [email protected]
Website: http://www.contecmed.com
Address: Eiffestrasse 80, 20537, Hamburg, Germany
Tel: +49-40-2513175
Fax: +49-40-255726
E-mail: [email protected]
RF communications equipment and pulse oximeter probe–
for pulse oximeter probe that are not LIFE-SUPPORTING
Recommended separation distances between
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2 P
0.12
0.37
1.17
3.69
11.67
CONTEC MEDICAL SYSTEMS CO., LTD
Shanghai International Holding Corp. GmbH (Europe)
pulse oximeter probe
electromagnetic interference by
(m)
80 MHz to 800 MHz
d = 1.2 P
0.12
0.37
1.17
3.69
11.67
File No.: CMS2.782.G001(CE)ESS/1.6
File Ver.: 1.6
Release Date: 2020.06
800 MHz to 2.5 GHz
d = 2.3 P
0.23
0.74
2.33
7.38
23.33
7