Apogee SI-4H1 Instrukcja obsługi - Strona 8

Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Przyrządy pomiarowe Apogee SI-4H1. Apogee SI-4H1 20 stron. Infrared radiometer
Również dla Apogee SI-4H1: Instrukcja obsługi (20 strony), Instrukcja obsługi (20 strony)

Apogee SI-4H1 Instrukcja obsługi
Field of View
The field of view (FOV) is reported as the half-angle of the apex of the cone formed by the target surface (cone
base) and the detector (cone apex), as shown below, where the target is defined as a circle from which 98 % of the
radiation detected by the radiometer is emitted.
Sensor FOV, distance to target, and sensor mounting angle in relation to the target will determine target area.
Different mounting geometries (distance and angle combinations) produce different target shapes and areas, as
shown below.
A simple FOV calculator for determining target dimensions based on infrared radiometer model, mounting height,
and mounting angle, is available on the Apogee website:
http://www.apogeeinstruments.com/using-your-apogee-
instruments-infrared-radiometer/.