3 Installation Procedure

3.1 Shimadzu LC-10/20 Pump communication

The Shimadzu LC- 10/20 Pump is controlled by serial (RS232)
communication. The communication is provided via the Optical to USB
convertor connected to the computer via USB A-A cable (p/n SK10) and to
the pump via optical cable (p/n SK15). Both of these cables are part of the
DataApex Opt-USB converter set (p/n OPT01), but can also be ordered
separately.
Optical to USB converter
For communication an OPT-RS232 converter is needed. It can either be
the DataApex Opt- USB converter (p/n OPT01) or Shimadzu adaptor
(Shimadzu p/n 22-35480-91).
Pump Settings
Communication parameters should be set on the pump in the SERVICE
MODE (see pump documentation) as follows:
Baud rate 2400, parity N, bits 8, stop bit 1.
SYS (Set System Parameter) has to be set to 5 for isocratic mode or to 6
for gradient mode from the pump keyboard.
Caution:
These values are not documented in the pump manual.
The other Pump system parameters should be set to:
Local=0 (Remote)
ADRS= any value (not used)
CBM LINK=0 (not used)
Shimadzu LC-20 settings
Shimadzu LC-20 pumps have to be set to the LC-10 emulation mode in
order to be able to work in Clarity. This is done from the pump's keyboard
according to this procedure:
Press the VP button on the pump's front keyboard until the CALIBRATION
l
inscription is shown on the pump's display.
Press the func button to enter the pump's configuration. The password
l
has to be filled in.
Note:
The default password is 00000.
Press the func button repeatedly until the OP MODE inscription is shown
l
on the pump's display.
Enter the desired number ( 1 ) to switch the pump to the SCL- 10Avp
l
emulation mode and press the enter button.
- 3 -
3 Installation Procedure