Durst LABORATOR 184 Instrukcja obsługi - Strona 16

Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Urządzenia przemysłowe Durst LABORATOR 184. Durst LABORATOR 184 18 stron. Counter-weight spring adjustment

1)
Head carriage in bottom position.
2)
Spring is tight.
3)
Install tension tool
4)
Turn to slack wire.
5)
Secure tension tool with "C" bracket.
6)
Remove wire disk in opposite side of tension tool.
7)
Slack spring until it is fully slack or blocked by wire taken up.
8)
Install nut and pin in side without wiredisk installed.
9)
Move tool to opposite side - the side with no wire disk and where
you just installed the nut and pin.
10)
Tighten slightly to allow the second wiredisk to be removed.
11)
after both wire disk are removed tension spring fully.
12)
Install the new wire in the wire disk
13)
Install wire wheel in opposite side of tension tool.
14)
Release the spring until spring is completely slack or blocked
by the wire taken up.
15)
move the tension tool to the side with the wire wheel installed.
16)
Tighten spring completely or until further motion is blocked by
the wire taken up
17)
thread wire through the head carriage loop and through the spring
housing.
18)
pull the wire tight while sliding last wire wheel into position.
Avoid kinks.
19)
install and secure wire in wire wheel.
20)
replace wire wheel covers.
end.