Display connections
Note: Communication
connector must be aligned
with socket. If out of
alignment, controller will not
boot properly.
14.
F
Release lower wire bundle. Locate wires labelled X3 and ID2 and connect as indicated in schematic
257349 near the bottom of the page. Schematic refers to 'existing components'. See original
chamber schematic to determine which option was installed. It will either be:
i.
A normally open contact on the contactor, or
ii.
A Chamber On Relay (COR).
15.
F
From the lower wire bundle, locate the ground wire. Attach ground wire ring connector to
threaded stud and secure with additional lock washer and nut (Figure 11). Do not to overtighten.
1
CONVIRON
2
3
4 5 6
Figure 10
1 - Control panel door
2 - Retrofit cover panel
3 - New control panel assembly
4 - Star washer and nut
5 - Ground wire ring connector
6 - Star washer and nut
Figure 11
CMP 6050 RETROFIT INSTALLATION INSTRUCTIONS
Buzzer wires
Wires for buzzer
connection routed
through opening in
control panel assembly.