This symbol means: "Consult operating instructions". To achieve accurate results, read
the instructions for use carefully before use and follow all instructions provided therein.
Keep the instructions for use for future reference.
Your ADE Electronic Wheelchair Scale is a class III approved scale which has been manufactured for
medical use in hospitals, by doctors or medical specialists as well as in rehabilitation centres or
nursing homes.
Portable and mobile HF-communication systems may influence electronic medical devices.
Electronic medical devices are restricted to special safety regulations in respect of their electromag-
netic compatibility and have been positioned and operated strictly as per the instructions of this Opera-
ting Manual.
The scale is used to evaluate the weight of patients sitting in a wheelchair, able to walk on their own or
by using any kind of walking aid or lying on an emergency litter. Due to its high capacity the scale is
also used for adipose patients.
The scale is a precision instrument for medical use.
The weighing result is used as an essential basis for medical application decided by doctors or medi-
cal personnel in hospitals, nursing homes, rehabilitation- and dialysis centres.
In order to achieve the best possible results kindly read and follow the Operating Manual in all re-
spects.

Check carton contents

The scales are already mounted trough and ready for use.
In the scope of supply are contained:
-
Platform Scale compline mounts
-
Indicator and cable with jag connector system
-
Mains adapter
-
Operating Manual including Certificate of Conformity
In any case, check that the serial numbers on the scales and the display unit match!
This is a mandatory precondition for using the scales in legal metrological applications.
Observe the safety instructions!

Assembling the scales

The ADE-MAK display unit can be mounted on walls or bases, or can be used as a stationary unit on
a table.
If your scales are not connected to the ADE-MAK, connect your scales to the ADE-MAK scales con-
nection point (Fig. 2-1) using an appropriate connecting cable.
You must make sure that the display unit is set up dry, without being knocked, and that it is secured
against falling. Connect the data cable for the display unit to the scales.
Unpack the scale and the display carefully.
Place a scale onto a solid and flat underground so that all feet have contact with the ground.
If the scale was transported or moved, the scale must be lined up again.
M501020-210322-Rev006-UM-e

Installation and use of scale

3