Pressure Regulated Perfusion System
AutoMate Scientifi c now offers a Pressurized Perfusion Upgrade
with either four or eight channels for:
• Faster Switching
• Microliter fi ne delivery
• Small diameter pipette delivery
For Single-cell or Whole-cell Superfusion
Multi-barrel Perfusion Pencil
8 micro-bore tubes
into 1 outfl ow
• Single cell and
patch delivery
• Rapid solution
change
• Micro-liter dead
volume
• 250 & 100 micron
removable tips
4
• Steady fl ow-rate
• Microinjecting
The package can be added to
any new or existing per-
fu sion rig - from AutoMate,
any third-party man u fac -
tur er, or even homemade.
Connect to house air, tank,
or compressor (30 to 100
psi). Precision regulator
delivers zero to 10 psi.
Perfusion System
1
2
3
4
vacuum
trap
*Unused lines should be replaced with
plugs on manifold, or tem po rari ly
clamped-off just above man i fold to
avoid back bleeding into empty lines.
Priming one line at a time reduces
bub bles.
Connect your vacuum trap line directly
to the outfl ow of the man i fold for run-
ning cleaning solutions and drying your
lines quickly.
Determine dead volume from the
manifold to sample by measuring
time and volume required to fl ush a
colored liquid with a clear one.

System Set-up Diagrams

Instructions on this page for Econ-
o my per fu sion systems, Lure-lock
fi ttings, Pres sur ized systems or Pinch
valve systems. (Regular hose barb
Tefl on perfusion systems please see
the next page.)
Lure-lock systems should also see
photo on page 3. Find additional
photos and diagrams of the pressure
system on pages 4 and 8.
Flow rate adjusted by relative height of
reservoirs and included fl ow regulator.
Always calibrate fl ow rates - regulator
values are relative (see below).
Leave tubing between reservoirs and
valves long enough to raise and lower
reservoirs.
1) Assemble ringstand.
2) Attach valve unit.
3) Attach reservoir bracket(s).
4) Insert reservoirs.
5) Attach provided tubing to upper
valve barb (cut as desired, but see
note above fi rst).
6) Connect tubing between bottom
valve barbs and manifold.
7) Connect fl ow regulator tubing
to manifold outfl ow and your
per fu sion chamber (downsizing if
necessary).
8) Plug numbered valve cables into
ValveLink ports.
Calibrate fl ow rates by running one line
into a graduated cylinder for a fi xed
pe ri od of time.