Troubleshooting:
Blank, featureless disk:
Thin clouds or haze may be
try again later.
Check solar activity, no
may be present.
Try moving focus in and out by 1-2 inches (2-5cm).
Ensure Quark is installed AFTER diagonal.
Make sure a 25 to 40mm eyepiece (not included) is installed.
Ensure knob is pointing straight away from the power
Check for focuser droop, all connections between telescope and
Quark must be tight and square.
Poor contrast;
Check that optical surfaces are clean. Dust
do not affect the
view, check for smudges such as oil from fingerprints.
Try adjusting wing shift knob up 3-4 clicks clockwise
center and
wait 5-10 minutes. If that does not improve the view,
try
down 3-4 clicks
below center and wait 5-10 minutes.
Atmospheric
or
may be poor,
again later.
image:
Blurry views are typically due to poor
Poor
can be
caused by the presence of heat waves from concrete, asphalt, or machinery.
Weather effects like the
stream can also cause blurring.
moving to a
different location or else observe on a day when weather conditions are
improved.
Yellow LED indication:
Your filter is adjusting to the wavelength chosen by the knob. Wait
approximately 5-10 minutes for the temperature to adjust and then light
should turn green, indicating that the filter has settled to its required
temperature and is on band for viewing.
Yellow, never goes green LED indication:
If after 20 minutes of the same knob position
the LED has
not turned green, the ambient
may be too hot or too cold for the
Quark to regulate the
However, the filter may still be usable
while slightly mistuned and performance may not be affected.
Red LED indication:
This usually indicates that the Quark is not receiving enough
voltage. If
by
the
Make sure to use the
supplied
AC/DC
wall adapter, as many cell phone chargers and PC USB
ports do not have
current
for the Quark.
Red indication can also mean an electrical fault in the Quark
electronics.
If power source changes do not resolve the red indication,
return the Quark to
for inspection and