Gait up Physilog 5 Podręcznik użytkownika - Strona 14

Przeglądaj online lub pobierz pdf Podręcznik użytkownika dla Czujniki bezpieczeństwa Gait up Physilog 5. Gait up Physilog 5 17 stron.
Również dla Gait up Physilog 5: Podręcznik użytkownika (12 strony), Podręcznik użytkownika (17 strony)

Gait up Physilog 5 Podręcznik użytkownika
longer waterproof. The forced hard reset is the ultimate way to make Physilog® responsive again, it
should only be used in desperate cases. Before you apply it, let the sensor rest for 5 minutes
(especially if the sensor is unresponsive due to a problem during a firmware update) and be sure
that the sensor is not unresponsive due to completely discharged battery.
LED indications
Situation
Description
Connected via USB
Green continuous
Orange blinking
Multicolor blinking
Red continuous
During
standalone
Green blinking
measurement
Green continuous
Orange blinking
During
Green blinking (2 flashes per blink)
measurement
Orange blinking (2 flashes per blink)
started from
application
Green/orange blinking (1 green, 1 orange flash
per blink)
Error indications
2 rapid red blinks
3 rapid red blinks
4 rapid red blinks
5 rapid red blinks
6 rapid red blinks
Troubleshoot for error indications
Error Type*
Troubleshoot
Error 1
Disconnect and replug the Physilog®
Error 2
Contact Gait Up
Error 3
Charge the Physilog®
Error 4
Contact Gait Up
Error 5
Contact Gait Up
Error 6
Empty the memory card
*see LED indication above
Physilog5 user manual v1.2.6 Date: 19.06.2018
Copyright © 2018, All Rights Reserved - Gait Up SA. CH-1015 Switzerland
EPFL Innovation Park, Bat C. CH-1015 Lausanne
Meaning
Fully charged
Charging
Data transfer
Error 1
Measuring
Waiting for synchronization
Low battery
Measuring and sending data to application
Measuring, connection to application to send
data lost
Measuring and sending data to application with
low battery
Error 2
Error 3
Error 4
Error 5
Error 6
Gait Up SA
www.gaitup.com
TVA CHE-145.136.760
14