STORAGE
1. Switch off power from main and unplug,
separate the pump from the mattress
and place the mattress flat.
2. Fold the mattress in two from mattress
foot to head.
3. Allow time for air to escape from
mattress to reduce storage space.
4. Close the protective cap cover on the
quick connector.
SPARE PARTS/ACCESSORIES
For spare parts and accessories please
contact Dynarex or an authorized dealer.
GENERAL CONDITIONS OF DISPOSAL
Dispose of the device in the appropriate
disposal areas for recycling.
At the end of its life, the product must not
be disposed of along with other domestic
waste. Dispose of this equipment by
bringing it to a designated recycling point
for electric and electronic equipment or
others that provide this service.

TROUBLESHOOTING

If your questions can't be answered with
below information, please contact your
local agent directly. They might require a
technician to take care of the problem.
No lights on unit
Low pressure alert
Patient's body is
sagging in the
middle/bottoming out
Control lock up
or "freeze"
Check control unit is connected to the main power supply and that
the unit is switched on.
Check quick blow fuse on the control unit or mains plug fuse.
Check main power to the control unit and that the plug is switched.
Check mattress air supply connector is connected to the pump
unit correctly.
Check CPR connector is connected properly.
Under the mattress cover check all "T" & "L" connectors are
connected to the air tubes.
Check the connector tubes for kinks, crimps or damage.
Increase the value of the comfort setting. Wait a couple of minutes
so pressure can stabilize before making another change.
Turn the control unit off and unplug the pump. Wait a few minutes
and plug the control unit back in. Turn on the control unit.

MAINTENANCE

In case of repair, use only original spare
parts and accessories.

CLEANING AND DISINFECTION

WARNING!
Unplug the pump before starting
the cleaning procedure.
1. Clean the mattress with mild soap.
2. Do not immerse or soak pump unit.
3. After cleaning, dry the mattress without
direct sunlight exposure.
4. Check that the connection tubes are not
obstructed by dust or other object. Use a
soft swab to clean the tubes if necessary.
NOTE:
Never use acids, alkalis or solvents
such as acetone to clean the mattress.
5