iHealth BP7 Instrukcja obsługi - Strona 11

Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Monitor ciśnienia krwi iHealth BP7. iHealth BP7 16 stron. Wrist
Również dla iHealth BP7: Instrukcja obsługi (16 strony), Skrócona instrukcja obsługi (4 strony)

iHealth BP7 Instrukcja obsługi
6. It is recommended that product performance be checked every 2 years or after each repair.
Please contact the service center.
7. No monitor component needs to be maintained by the user. The circuit diagrams, component
part lists, descriptions, calibration instructions, or other information which will assist the user's
appropriately qualified technical personnel to repair those parts of the equipment which are
designated for repair can be supplied.
8. Clean the monitor with a dry, soft cloth or a moistened and well wrung soft cloth using water,
diluted disinfectant alcohol, or diluted detergent
9. The monitor can maintain the safety and performance characteristics for a minimum of 10,000
measurements or three years of usage.
10. The battery can maintain the performance characteristics for a minimum of 300 charge cycles.
Battery replacement should only be performed by a qualified iHealth technician. To do
otherwise will void your warranty and possibly damage your unit.
11. Cuff replacement should only be performed by a qualified iHealth technician. To do otherwise
will possibly damage your unit.
12. It is recommended that if the cuff is used, for example, in a hospital or a clinic, it be disinfected
twice a week. Wipe the inner side (the side that contacts skin) of the cuff with a soft cloth
lightly moistened with Ethyl alcohol (75-90%). Then air dry the cuff.

WARRANTY INFORMATION

The iHealth Wireless Blood Pressure Wrist Monitor is warranted to be free from defects in
materials and workmanship within one year from the date of purchase when used in accordance
with the provided instructions. The warranty extends only to the end user. We will, at our option,
repair or replace without charge the iHealth Wireless Blood Pressure Wrist Monitor covered by the
warranty. Repair or replacement is our only responsibility and your only remedy under the
warranty.

EXPLANATION OF SYMBOLS

Symbol for "THE OPERATION GUIDE MUST BE READ"
The sign background color: blue
The sign graphical symbol: white
Symbol for "WARNING"
Symbol for "TYPE BF APPLIED PARTS" (Cuff only)
Symbol for "ENVIRONMENT PROTECTION – Waste electrical products should not be
disposed of with household waste. Please recycle where facilities exist. Check with your
local authority or retailer for recycling advice".
Symbol for "MANUFACTURER"
Symbol for "COMPILES WITH MDD93/42/EEC REQUIREMENTS"
10