Emerson Micro Motion TA Series Instrukcja instalacji - Strona 11
Przeglądaj online lub pobierz pdf Instrukcja instalacji dla Przyrządy pomiarowe Emerson Micro Motion TA Series. Emerson Micro Motion TA Series 18 stron. Flow and density meter
Również dla Emerson Micro Motion TA Series: Instrukcja instalacji (20 strony), Instrukcja instalacji (16 strony)

Installation Manual
MMI-20040393
3
Wiring
3.1
Tantalum wiring
•
A 9-wire cable is limited to 1,000 ft (304.80 m) between the sensor and the
transmitter.
•
If you are using the Smart Meter Verification (SMV) application, the 9-wire cable is
limited to 60 ft (18.29 m) maximum.
•
Use Teflon
Make sure the hazardous area specified on the sensor approval tag is suitable for the
environment in which the sensor will be installed. Failure to comply with the
requirements for intrinsic safety in a hazardous area could result in an explosion.
Fully close and tighten all housing covers and conduit openings. Improperly sealed
housings can expose electronics to moisture, which can cause measurement error or
flowmeter failure. Inspect and grease all gaskets and O-rings.
3.1.1
Connect the 9-wire cable
Procedure
1. Prepare and install the cable according to the instructions in the Micro Motion 9-Wire
2. Insert the stripped ends of the individual wires into the terminal blocks of the
Installation Manual
™
jacketed cable for SMV applications.
DANGER
CAUTION
Flowmeter Cable Preparation and Installation Guide.
junction box. Ensure that no bare wires remain exposed.
Terminal block number
1
Wire color
Blue
Wiring
April 2019
11