2M Technology 2MTHWT-HMD Skrócona instrukcja obsługi - Strona 3

Przeglądaj online lub pobierz pdf Skrócona instrukcja obsługi dla Czujniki bezpieczeństwa 2M Technology 2MTHWT-HMD. 2M Technology 2MTHWT-HMD 12 stron. Walkthrough metal detector
Również dla 2M Technology 2MTHWT-HMD: Podręcznik użytkownika (20 strony), Podręcznik użytkownika (8 strony)

2M Technology 2MTHWT-HMD Skrócona instrukcja obsługi
Disclaimer
The infrared temperature measuring device is the human body surface temperature
screening instrument. The test results only reflect the real-time temperature evaluation
of objects or people in a specific environment. The selling company and the manufacturer
shall not bear any responsibility for the direct or indirect loss caused by the use of the test
results.

Installation precaution

1. It is recommended to install it indoor free from wind. The cold wind might influence the
accuracy in windy outdoor. In this case, we have to increase the environment temperature
compensation, 5° C per time.
2. The temperature measurement area should be protected from sunlight to avoid high results.
3. When it is used outdoor under special circumstances, build a shed on site. The height of
the shed should be more than 2.4m (7.87ft) to ensure the stability and free from wind.
People walk slowly through it to measure the temperature.
Temperature measurement module
Temperature sensor:
Real time infrared thermo sensor to detect temperature,
ready in 30 seconds after power on.
Distance sensor:
The distance sensor starts working when a person or
object is in 20cm (7.87" in) to it. If the temperature of person
or object detected is ≥ 35.5 ℃, the voice "Normal
temperature" is trigged; if it's lower than 35.5 ℃, or higher
than the set value with alarm, the voice "Please check
again" is trigged.
When the ambient temperature is lower than 15 ℃, if the
temperature of person or object detected is ≥ 34.5 ℃, the voice
"Normal temperature" is trigged; if it's lower than 34.5 ℃, or higher
than the set value with alarm, the voice "Please check again" is
trigged.
Special attention: the distance sensor shall not be blocked,
otherwise it will repeat the voice broadcast.
Power Light:
The indicator light is on when the temperature measurement
module is working. When measuring the forehead temperature,
the measured result is the most accurate when the middle of
human eyes is right facing the temperature indicator.
Body Temperature Measurement – Quick Guide V 6.1
Temperature Measurement
Power Light
Power Light
Distance sensor
Temperature sensor
Sketch Map