2M Technology 2MTHWT-HMD Skrócona instrukcja obsługi - Strona 8

Przeglądaj online lub pobierz pdf Skrócona instrukcja obsługi dla Czujniki bezpieczeństwa 2M Technology 2MTHWT-HMD. 2M Technology 2MTHWT-HMD 12 stron. Walkthrough metal detector
Również dla 2M Technology 2MTHWT-HMD: Podręcznik użytkownika (20 strony), Podręcznik użytkownika (8 strony)

2M Technology 2MTHWT-HMD Skrócona instrukcja obsługi
● L001-L006 of the first line is the metal detection zone from bottom to top, 975 (or other values) of the
second in represents the metal detection sensitivity, high sensitivity for tiny metals.
● D001 of the first line is the alarm time setting, and 1.0 of the second line means the alarm time is 1 second.
36. 5
L 006
......
975
Metal detection on setting
Press D button in the main interface (display CCCC), press B button to enter the password input interface
(display 1234), and press B button to shift and C button to adjust to input the password 1111. Press D button
to display FFFF, indicating the password is correct. Press B button to enter the function setting interface
(display L001/975). Press B button to shift to L 007 / 0 , and press B button to shift and C button to adjust to
change 0 to 975.Press B button to shift to d001 option, and press D button to complete the setting (display
0000). Press A button twice to return to the main interface.
● L007 of the first line represents all the 6 metal detection zones. The number 975 of the second line
represents the metal detection sensitivity.
● L008 of the first line is the setting to detect small size metals. The number 1 of the second line represents
to detect small size metals, and 50 means not detect small size metals.
● D001 of the first line is the alarm time setting, and 1.0 of the second line means the alarm time is 1 second.
36. 5
L 007
d001
975
1. 0
Instruction of temperature measurement
1. Power on and wait for 30 seconds, keep nothing in front of the temperature sensor.
2. Normal ambient temperature is 0°C - 35°C. When using it at - 10°C, you have to decrease the
environment temperature compensation, 5°C per time, until the result difference between it and the
forehead thermometer is less than ± 0.3°C.
3. Measurement method: Measure forehead or wrist surface
4. Forehead temperature and wrist surface temperature can be measured without adjusting settings.
5. Measurement area: the center of forehead or the wrist covered by sleeve.
6. Standard measurement distance: 1cm-10cm (0.39"-3.93" in). When measuring at a distance of 10-30cm
(3.93"- 11.81" in), it is necessary to adjust the environment temperature compensation, 5°C per time,
until the result difference between it and the forehead thermometer is less than ± 0.3°C.
1234
1111
0000
0000
L 006
d001
1. 0
1234
1111
0000
0000
0000
0000
Body Temperature Measurement – Quick Guide V 6.1
FFFF
L 001
0000
975
0000
0000
FFFF
L 001
0000
975
36. 5
L 001
36. 5
L 007