Avante Volumax Skrócona instrukcja obsługi - Strona 4
Przeglądaj online lub pobierz pdf Skrócona instrukcja obsługi dla Sprzęt medyczny Avante Volumax. Avante Volumax 4 stron. Vos
ALARM ICON
CARTRIDGE FAULT
CARTRIDGE TYPE
GAS SUPPLY
(FLASHING)
GAS SUPPLY
(CONTINUOUS AND FLOW
RATE NUMERIC DISPLAY
FLASHES)
GENERAL FAULT
TEMPERATURE DISPLAY
SHOWS DASHES (
FLASHING
TEMPERATURE NUMERIC
DISPLAY FLASHES
2601 Stanley Gault Pkwy, Suite 101
800-923-8413
502-244-6345
•
INDICATES
Cartridge and/or
DPC not detected.
Unit will not run.
Gas bubbles in water
circulation. Unit
continues to operate.
Cartridge and/or
DPC not detected.
LO
HI
Indicates type of cartridge installed (low or high flow). Not an alarm.
Gas supply pressure
outside 4-85 psi
(28-586 kPa) range.
Unit will not operate.
Selected flow can
not be provided from
current gas supply.
Temperature out
of range
- -
)
Temperature 2° >
set point
Temperature 2° <
set point
GENERAL FAULT ALARMS:
oxygen percentage will cause a General Fault alarm indicated by this icon. Unit will continue gas delivery in a fault
condition until placed in Standby. With the exception of O2 sensor replacement, the unit must be repaired by an
approved service facility. General Fault alarms cannot be silenced with the mute button. To reset, first disconnect
the unit from AC power and then press the Run/Standby Button.
Louisville, KY 40223 USA
•
FAX: 502-244-0369
•
CAUSE
RUN mode: faulty
Disconnect patient. Remove disposable
sensor or cartridge
patient circuit. Check cartridge installation.
not detected.
Check sensor windows are clean
Excessive gas
Disconnect patient. Shut off unit. Replace
diffusion through
disposable patient circuit including water
cartridge fibers.
path, cartridge & delivery tube.
STANDBY mode:
Remove disposable patient circuit.
missing cartridge.
Check cartridge installation.
Gas supply is
disconnected or
Check gas supply and correct as necessary.
exhausted.
Inlet gas pressure too
Increase gas pressure or decrease flow
low for selected flow
setting.
rate.
Overheating or
Cannot be corrected by user: disconnect
temperature sensor
patient. Shut off unit and send for service.
malfunction.
User enters set point
Silence alarm and wait for
much lower than
temperature to drop.
previous temperature.
Very low water
Silence alarm and wait for
temperature after bag
temperature to rise.
replacement.
Failures in the control or measurement systems for temperature, gas flow, and
www.avantevet.com
ACTION