Ceyear 4024D Podręcznik konserwacji - Strona 19

Przeglądaj online lub pobierz pdf Podręcznik konserwacji dla Przyrządy pomiarowe Ceyear 4024D. Ceyear 4024D 19 stron. Spectrum analyzer

Ceyear 4024D Podręcznik konserwacji
through 【System】-[Network Configuration].
Attention: The IP address of the 4024 spectrometer and PC shall be located on the same network
segment. If they are different, the IP of the computer or spectrometer shall be set to make it within the
same network segment (for example, if the IP of the spectrometer is 172.141.11.202, the IP of the
computer shall be set to 172.141.11.xx).
Click the indexes to be tested, and click Start Test.
Attention:
① In case of measurement and display of the average noise level and remaining response, a 50Ω load
shall be connected at the RF port. The indicators including test frequency range, third-order intercept
point, 1 dB gain compression and standing-wave ratio can only be manually tested.
② In case of initial measurement of the reference level, scale fidelity and total level uncertainty, the
power meter shall be connected for calibration. No recalibration is needed when another spectrometer is
tested.
4)
After the test of all indexes is completed, click Save Data to save the test results in the form of word
files.