Level accuracy
(25℃± 10℃)
Output impedance
SWR
Maximum reverse power
Spectral purity(specification is point frequency without modulated mode)
Harmonics
(Measured at +5dBm or
maximum specified power,
whichever is lower)
Non-harmonics
(0dBm,>10kHz offset)
SSB Phase noise(at
maximum stable output
power)
40GHz<f≤50GHz
-10dBm<P≤Maximum
stable output power
-60dBm<P≤-10dBm
-90dBm<P≤-60dBm
50Ω(Rating)
1MHz≤f≤20GHz
20GHz<f≤40GHz
40GHz<f≤50GHz
+27dBm(0V DC) (Rating)
1MHz≤f≤1.5GHz
1.5GHz<f≤2.5GHz
2.5GHz<f≤19GHz
19GHz<f≤25GHz
25GHz<f≤50GHz
1MHz≤f<2.5GHz
2.5GHz≤f<5GHz
5GHz≤f≤10GHz
10GHz<f≤20GHz
20GHz<f≤38GHz
38GHz<f≤50GHz
1MHz≤f<2.35GHz
2.35GHz≤f<2.5GHz
2.5GHz≤f<5GHz
5GHz≤f≤10GHz
-120dBm~0dBm
± 1.0dB
± 1.5dB
± 1.8dB
<1.8:1
<2.0:1
<2.5:1
≤-40dBc
≤-30dBc
≤-40dBc
≤-30dBc
≤-35dBc(typical)
≤-54dBc
≤-60dBc
≤-56dBc
≤-50dBc
≤-44dBc
≤-40dBc
≤-82dBc/Hz@100Hz
≤-98dBc/Hz@1kHz
≤-108dBc/Hz@10kHz
≤-106dBc/Hz@100kHz
≤-94dBc/Hz@100Hz
≤-110dBc/Hz@1kHz
≤-120dBc/Hz@10kHz
≤-118dBc/Hz@100kHz
≤-88dBc/Hz@100Hz
≤-104dBc/Hz@1kHz
≤-114dBc/Hz@10kHz
≤-112dBc/Hz@100kHz
≤-82dBc/Hz@100Hz
≤-98dBc/Hz@1kHz
≤-108dBc/Hz@10kHz
≤-106dBc/Hz@100kHz
5