TROUBLE SHOOTING
If you experience difficulties in operating your AliveCor products, refer
to the troubleshooting guide below or contact technical support at
Problem:
My KardiaBand is not working.
I have a lot of artifact, noise, or
interference in my recording, or "No
EKG recorded" message displays.
The ECG rhythms appear
upside down.
Solution:
Option 1: Ensure that the Kardia
watch app has access to the watch's
microphone. On the iPhone, go to
Settings and tap the Kardia app.
Tap the microphone toggle.
Option 2: Ensure that the watch
microphone is unobstructed.
Consult the watch user manual
if it is obstructed.
Option 3: Check and change the
battery. Use a Torx T3 screwdriver
for the sensor battery door.
Option 1: Ensure that your watch,
arms, and hands remain still during
recordings.
Option 2: Clean the electrodes on
the KardiaBand with an alcohol-
based sanitizer.
Option 3: If your hands are very dry,
use a water-based lotion before
recording.
Option 4: When recording, relax your
arms and hands to reduce muscle
noise. Rest the forearms and hands
on a flat surface.
Option 1: The watch orientation may
be set to the wrong wrist. On your
iPhone, go to the Watch app. Tap My
Watch > General > Watch Orientation.
Option 2: The KardiaBand pieces
may be attached to the watch in the
wrong orientation. Review "Assembly"
instructions.
CAUTIONS
General:
DO NOT store in extremely hot, cold,
humid, wet, or bright conditions.
DO NOT expose to strong electro-
magnetic fields.
DO NOT take recordings in close
vicinity to other equipment emitting
ultrasonic acoustics.
DO keep components out of reach
of children.
DO use this device to record heart rate
and heart rhythm only.
DO NOT use the sensor on portion
of the body with too much body fat,
body hair or very dry skin, a successful
recording may not be possible.
AliveCor makes no warranty for any
data or information that is collected
erroneously by the device, or misuse
or malfunction as a result of abuse,
accidents, alteration, misuse, neglect,
or failure to maintain the products as
instructed. Interpretations made by
this device are potential findings, not
a complete diagnosis of cardiac
conditions. All interpretations should
be reviewed by a medical professional
for clinical decision-making.
KardiaBand:
DO NOT use with a cardiac pacemak-
er, ICDs, or other implanted electronic
devices.
DO NOT continue use until further
instructed by a physician if your skin is
irritated or inflamed around the sensor
or band.
DO NOT drop or bump with excessive
force.
DO NOT use to diagnose heart-
related conditions.
DO NOT expose the device to a mag-
netic resonance (MR) environment.
DO NOT wear during cautery and
external defibrillation procedures.
After ECG analysis, the app may
incorrectly identify ventricular flutter,
ventricular bigeminy, and ventricular
trigeminy heart conditions as
unreadable. Please consult with
your physician.
CAUTION: AliveCor does not guar-
antee that you are not experiencing an
arrhythmia or other health conditions
when labeling an ECG as normal. You
should notify your physician for possi-
ble changes in your health.
KARDIABAND
SPECIFICATIONS
Battery: CR1620 Coin Cell
Storage Conditions: Original
package under normal room
temperature and humidity
The KardiaBand sensor may be
removed and inserted a maximum
of 50 times without performance
degradation.
ELECTROMAGNETIC & OTHER
INTERFERENCES
The KardiaBand has been tested
and deemed in conformance with
the relevant requirements in
IEC60601-1-2:2014 Class BF for
Electromagnetic Compatibility (EMC).
INGRESS PROTECTION
MARKING
KardiaBand is IP64 rated. Kardia-
Band is protected against insertion of
fingers and is not affected by vertically
dripping water. KardiaBand has been
tested with relevant requirement stan-
dard IEC60601-1-11:2015.
EQUIPMENT SYMBOLS
Serial number
Model number
European Authorized
Representative
Manufacturer
Read instructions
before use
Type BF applied part
Temperature range
Humidity range
Do not dispose with
household waste
ADDITIONAL INFORMATION
For more detailed troubleshooting
and technical information, please visit:
https://www.alivecor.com/
support/#user-manual