Zero Tracking Setting [ trAcE ]
Zero Tracking is a function to help compensate the weight readout fluctuations at zero point to
improve the usability of the scale. The Acaia Orion is allowed to compensate fluctuations by up to
2d according to industrial weighing guidelines.
1. OFF: It is not recommended to turn off Zero Tracking as the zero indication will drift easily as
the room temperature changes.
2. 0.5 d~3 d: For example, since each division is 0.1 g, 3 d = 0.3 g. When Zero Tracking is set to
be 3 d, any initial weight between +/- 0.3 g on the weighing pan will be zeroed out.
3. 2 d is recommended in general for the Orion.
4. To save your selection and exit the settings sub-menu, press Dose. [ trAcE ] will display on
the LED to confirm that the selected option has been saved.
Key Sound Setting [ bEEP ]
1. Set the Key Sound function to Off/On to turn off/turn on the beeping sound.
2. To save your selection and exit the settings sub-menu, press Dose. [ bEEP ] will display on the
LED to confirm that the selected option has been saved.
Default Weighing Unit [ Unit ]
1. The Default Weighing Unit is set to grams if [ 9 ] is selected and ounce if [ o2 ] is selected.
2. Rotate Dial to switch the Default Weighing Unit between grams and ounces.
3. To save your selection and exit the settings sub-menu, press Dose. [ Unit ] will display on the
LED to confirm that the selected option has been saved.
Calibration [ CAL ]
Please go to "Chapter 5: Calibration" for more information
About [ About ]
Shows the current Orion Firmware version.
To update to the latest version, please use Orion Updater on
www.acaia.co/pages/apps