18

Approximate Settings Based on Patient Height

2 Second Inspiratory Time
Tidal Volume
Height Male
=((IBW-50)/2.3)+60
inch
@10 ml/Kg
(cm)
Height Female
=((IBW-45.50)/2.3)+60
inch
@10 ml/Kg
(cm)
Flow Rate
8
BPM = I:E (1 TO - )
9
BPM = I:E (1 TO - )
10
BPM = I:E (1 TO - )
11
BPM = I:E (1 TO - )
12
BPM = I:E (1 TO - )
13
BPM = I:E (1 TO - )
14
BPM = I:E (1 TO - )
15
BPM = I:E (1 TO - )
16
BPM = I:E (1 TO - )
17
BPM = I:E (1 TO - )
18
BPM = I:E (1 TO - )
19
BPM = I:E (1 TO - )
20
BPM = I:E (1 TO - )
Ideal Body Weight Kg
@10 ml/Kg
400
480
560
56
59
63
4' 8"
4' 11"
5' 3"
(142.2)
(149.9)
(160.0)
58
61
65
4' 10
5' 1"
5' 5"
(147.3)
(154.9)
(165.1)
12
14.4
16.8
2.8
2.8
2.8
2.3
2.3
2.3
2.0
2.0
2.0
1.7
1.7
1.7
1.5
1.5
1.5
1.3
1.3
1.3
1.1
1.1
1.1
1.0
1.0
1.0
0.9
0.9
0.9
0.8
0.8
0.8
0.7
0.7
0.7
0.6
0.6
0.6
0.5
0.5
0.5
40.0
48.0
56.0
Flow Rates and I:E Ratios
By Tidal Volume and Breaths Per
Minute
640
720
800
66
70
73
5' 6"
5' 10"
6' 1"
(167.6)
(177.8)
(185.4)
68
72
75
5' 8"
6' 0"
6' 5"
(172.7)
(182.9)
(190.5)
19.2
21.6
24
2.8
2.8
2.8
2.3
2.3
2.3
2.0
2.0
2.0
1.7
1.7
1.7
1.5
1.5
1.5
1.3
1.3
1.3
1.1
1.1
1.1
1.0
1.0
1.0
0.9
0.9
0.9
0.8
0.8
0.8
0.7
0.7
0.7
0.6
0.6
0.6
0.5
0.5
0.5
64.0
72.0
80.0
17
880
960
1040
77
80
83
6' 5 "
6' 8"
6'11"
(195.6)
(203.2)
(210.8)
78
82
85
7'1"
6' 8"
6' 10"
(198.1)
(208.3)
(215.9)
26.4
28.8
31.2
2.8
2.8
2.8
2.3
2.3
2.3
2.0
2.0
2.0
1.7
1.7
1.7
1.5
1.5
1.5
1.3
1.3
1.3
1.1
1.1
1.1
1.0
1.0
1.0
0.9
0.9
0.9
0.8
0.8
0.8
0.7
0.7
0.7
0.6
0.6
0.6
0.5
0.5
0.5
88.0
96.0
104.0
1120
1200
87
90
7'3"
7'6"
(220.9)
(228.6)
89
92
7'5"
7'8"
(226.1)
(233.7)
33.6
36
2.8
2.8
2.3
2.3
2.0
2.0
1.7
1.7
1.5
1.5
1.3
1.3
1.1
1.1
1.0
1.0
0.9
0.9
0.8
0.8
0.7
0.7
0.6
0.6
0.5
0.5
112.0
120.0