CPS VG200 Instrukcja obsługi - Strona 4

Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Przyrządy pomiarowe CPS VG200. CPS VG200 4 stron. Digital vacuum gauge

CPS VG200 Instrukcja obsługi
The VG200 incorporates an easy to clean thermistor
sensor, which cannot be damaged from exposure to
positive pressure or oil mist. To clean the VG-200's
thermistor sensor, simply invert the unit and pour
approximately 1 tablespoon of 100% DENATURED
ALCOHOL directly into the sensor housing. Gently shake
the unit to produce a swirling effect of the cleaning
solution in the housing. Then, pour the remaining cleaning
solution out. Repeat this procedure until the alcohol
appears to be clean and clear of oil when removed from
the sensor housing. Then, allow the sensor to air dry (DO
NOT BLOW DRY). Check the gasket located inside the
nut during sensor maintenance. If the gasket appears to be
torn or deformed, replace it with a new gasket (CPS
P/N:HXG, includes 10 replacements).
SENSOR:
TYPE:
Self-heated thermistor bridge with integral temperature compensation
from 0 to 50 °C (32 to 122 °F).
OPERATING PRESSURE RANGE:
Maximum Working pressure: 600 PSIG
BURST PRESSURE:
OPERATING RANGE:
ACCURACY:
+/- 10% of Reading.
OPERATING TEMPERATURE RANGE:
0°C TO 50°C (32°F TO 122°F) Non-compensated (add +/- 0.5% error for every
°C outside compensated range): 0 to 50°C (-22 to 158°F)
OPERATING & STORAGE HUMIDITY:
STORAGE TEMPERATURE:
POWER SOURCE:
BATTERY LIFE:
minutes of operation.
WEIGHT:
153 g. (5.5 oz.) excluding the battery.
DIMENSIONS:
68 mm. diameter by 40 mm. deep by 124 mm. overall height
(2.7" dia. by 1.6" deep by 5" overall height).
MECHANICAL CONNECTION:
with core depressor.
BATTERY INDICATOR:

SPECIFICATIONS

3000 PSIG
Atmospheric to 0 microns
-40 to 85°C (-40 to 185°F)
One 9 volt alkaline battery (ANSI 1604AC – IEC 6LR61)
20 hours of continuous use. Instrument shuts itself off after 5
Standard 1/4" female SAE refrigerant hose type
Battery symbol with 3 segment Power Level Bar Graph.
Atmospheric pressure to 0 microns
Compensated (accuracy as stated):
0-95%, non-condensing.