AW Technologies TrachFlush Skrócona instrukcja obsługi - Strona 3
Przeglądaj online lub pobierz pdf Skrócona instrukcja obsługi dla Sprzęt medyczny AW Technologies TrachFlush. AW Technologies TrachFlush 4 stron.
TrachFlush Quick Guide
How to use Flush Control and Spont Flush Control
Pressure CONTROL settings
To perform a Flush Control, the
following settings are required.
(
Only adjust if clinically safe for the patient)
Pressure above PEEP
≥ 10 cmH
PEEP
≥ 5 cmH
Inspiratory time
≥ 1,33 s
OPTIONAL: If clinically safe for the patient, repeat Flush Control
Perform Flush Control
STEP 1: Press the Flush button to
activate Flush Control
STEP 2: When Flush Control is
complete, press
to return to main
screen
Pressure SUPPORT settings
To perform a Spont Flush Control, the
following settings are required.
(
Only adjust if clinically safe for the patient)
Pressure above PEEP
O
2
O
PEEP
2
Frequency
OPTIONAL: If clinically safe for the patient, repeat Spont Flush
Perform SPONT Flush
STEP 1: Hold the Flush button for 3
seconds. Read instructions, and press
"ACCEPT" to activate Spont Flush
STEP 2: When Spont Flush is
complete, press
screen
≥ 10 cmH
O
2
≥ 5 cmH
O
2
≤ 28 b/min
3 sec
to return to main
Page 3 of 4
Troubleshooting
STEP 1: If ventilator settings are
incorrect, a Flush Control is not
possible. Inspect the screen and
identify incorrect setting(s)
STEP 2: If clinically safe for the patient,
adjust the ventilator setting(s).
(See sections: Pressure Control/Support settings)
STEP 3: When correct, the Flush
Control automatically starts. Spont
Flush must be initiated manually