Electrical connection (3 x 20 A)
Voltage (default)
Rated load
Fuse rating
Minimum cross-section
Length of electrical cord (UL 921) with plug
Electrical connection
Voltage (convertible)
Rated load
Fuse rating
Minimum cross-section
Length of electrical cord (UL 921) with plug
Electrical connection
Cold water connections* (KW+DK)
Max. temperature
Max. water hardness
Recommended flow pressure
Min. flow pressure
Max. pressure
Throughput
Connection thread - on site (flat sealing)
Length of connection hose (supplied)
Length of the steam condensor hose
* Two cold water connections required: use supplied Y-adapter if only one cold water connection is available.
Hot water connection (WW)
Max. temperature
Max. water hardness
Recommended flow pressure
Min. flow pressure
Max. pressure
Throughput
Connection thread - on site (flat sealing)
Length of connection hose (supplied)
 
Installation plan PG 8582 CD / PG 8583 CD 
 
 
3 AC 208V 60 Hz
6.2 kW
3 × 20 A
4 × AWG 12
approx. 5'9"
NEMA L15-20
2 AC 208V 60 Hz
6.2 kW
2 × 30 A
3 × AWG 10
approx. 5'9"
NEMA L6-30
68 °F
73 gpg
29 psi
14.5 psi
145 psi
2 gal/min
3/4" Male Garden Hose
Thread
(USA: 11.5 NH)
approx. 5' 7"
approx. 5' 7"
150 °F
73 gpg
29 psi
5.8 psi
145 psi
2 gal/min
3/4" Male Garden Hose
Thread
(USA: 11.5 NH)
approx. 5' 7"
3 AC 208V 60 Hz
6.2 kW
3 × 20 A
4 × AWG 12
approx. 1.8 m
NEMA L15-20
2 AC 208V 60 Hz
6.2 kW
2 × 30 A
3 × AWG 10
approx. 1.8 m
NEMA L6-30
20 °C
70 °d
200 kPa
100 kPa
1,000 kPa
7.5 l/min
3/4" Male Garden
Hose Thread
approx. 1.7 m
approx. 1.7 m
65 °C
70 °dH
200 kPa
40 kPa
1,000 kPa
7.5 l/min
3/4" Male Garden
Hose Thread
approx. 1.7 m