Professional portable medical device for secretion removal, suitable for fixed and transportable use.
Complete with flame retardant ABS V0 casing and piston-cylinder electrical compressor lubricated for life.
Risk class under Directive 93/42/EEC:
Suction class:
Protection level against entry of liquids and solids:
Voltage:
Absorption:
Adjustable vacuum level:
Vacuum gauge precision class:
Air flow:
Temporary use:
Dimensions:
Weight:
Noise level:
Multi-voltage switching battery charger:
Internal battery:
Battery charge duration:
Operating conditions:
Temperature:
Air humidity:
Storage conditions:
Temperature:
Air humidity:
Operating-storage conditions:
Type BF appliance
It is compulsory to carefully read the instructions
before using this device
The device's casing is protected against
solid particles having diameter of 1 mm
or above, spraying water and access to
hazardous parts with a wire
Switch On
Switch off (or battery on charge)
Direct current
TECHNICAL CHARACTERISTICS
SYMBOLS USED
22
IIb
High vacuum / High flow
IP44
12VDC
3A
from 0 to –0.85 bar (-85KPa)
2.5% under UNI EN 837
30 lt/min operating without restrictions +/– 10 %
max. 45 minutes
38 (length) x 13.5 (width) x 26 (Height) cm
4.6 kg approx
55dBA (1m)
PRI: 100 ÷ 240V~ 50 – 60Hz
SEC: 14VDC 4.28A
12VDC 4Ah Lead, hermetic
45 minutes at maximum suction power
min. 0° C; max 40° C
min. 10 %; max 95 %
min. –10° C; max 50° C
min. 10 %; max 95 %
min. 690 hPa; max 1060 hPa
Alternate current
Do not use the unit when takig a bath or a
shower
Conforms to Directive 93/42/EEC for
medical devices
The device contains a hermetic lead
battery. It must be disposed of in
accordance with current regulations
on the disposal of toxic-harmful
waste
Sterilisation by ethyl oxide
Disposable
%