Memory:

Although the power is removed, the counter will memorize the last calculated number and average
size for each channel. This is saved until you perform a new counting operation on the particular
channel.
(Example: The power accidentally disappears when you are performing a counting operation. When
the power is applied again you can continue the counting process from where it left; just press the
Enter key when the screen reads WEIGHT(g) Cx: (x may be 1,2,3 or 4)).
The counter will remember the 15 latest counting operations, see User's Guide.

Previous average:

If the display reads WEIGHT(g) Cx: (x may be 1,2,3 or 4) and you only push the Enter key,
counting based on the previous average size can be done. This option must only be used when fish in
both the first and the second group are of the same size (cage/tank). If you use the previous average
size you will get the message StartCx: (number) Y/N. You are now able to start or continue the
counting process from the number of fish calculated during the previous counting operation.

Details about the rub window alarm:

If the screen reads RUB WINDOW C2, no counting will be performed on channel 2 and you will
loose the number of fish passed when this message appeared on the screen. The counting will be
normal in the other channels (if there is no other RUB WINDOW messages). It is not necessary to
stop the counter. If you press the C2 button the system will check the channel and if it is OK start to
count from the last number calculated just before the RUB WINDOW message appeared.
Note: AquaScan performs special filtration algorithms to "take care of" small particles, water etc. If
you get a RUB WINDOW message, be quick to push the corresponding key for the channel.
AquaScan will check the channel and continue counting unless something is still blocking the view of
the lenses.
General: During the counting, open the inspection-hatch regularly to check that the pipe/light source is
clean and look for droplets on the window in front of the lenses (the lenses may be disturbed even if
there have been no RUB WINDOW message). Also assure that the fish has a steady flow through the
unit.
Page 16 of 20