Troubleshooting and Maintenance
If you have any questions or concerns about the Vector Base System, please contact Bioness at
(800) 211-9136, Option 3 or contact your local distributor.
Troubleshooting
Issue
Trolley Squeaking
Harness Will Not Fit Patient Properly
Rope is Worn, Frayed or Damaged
Maintenance
Routine maintenance is critical for the safety and effectiveness of this device.
Check the following items before each use of the Vector Base System:
•
Inspect the trolley rope and rope knots. If the rope is worn, frayed or damaged contact
Bioness at (800) 211-9136, option 3 (in the United States) or your local distributor (outside
of the United States).
•
Inspect the patient harness for tears, material separation and Velcro degradation.
•
Inspect the Spreader Bar before each use.
•
Inspect the Spreader Bar rope and rope knots before each use.
•
Inspect the carabiners before each use.
Inspect the Mechanical Track quarterly. Clean debris and/or dirt when necessary. Tighten loose bolts,
screws or hardware if appropriate.
Items to be inspected and maintained on an annual basis:
•
Visually inspect the trolley wheels for wear.
•
Measure that the trolley spring is fully compressed when loaded with 180 lbs (81.6 kg).
•
Replace the trolley rope connecting the trolley to the Spreader Bar annually. Contact
Bioness or your local distributor to order a replacement part.
•
Replace the rope pulleys with attached carabiners annually. Contact Bioness or your local
distributor to order a replacement part.
Solution
Inspect track and drive wheels for debris and
dirt build up. Wipe off with a dry cloth.
• Verify that appropriate size is being used.
• If appropriate use a larger harness.
• Inspect Velcro and material to ensure integrity
is still intact.
• Replace rope.
• Call Bioness or your local distributor to order
a replacement part.
Chapter 5 - Troubleshooting and Maintenance