Acculab VICON Instrukcja obsługi - Strona 10

Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Wagi Acculab VICON. Acculab VICON 16 stron. Electronic precision scales

Acculab VICON Instrukcja obsługi

Totalizing

Purpose
With this application program you can add up successive weight values exerding capacity of balance/scale.
Example: Totalizing weight values
Step
1. Select application program
2. Select Totalizing
3. Confirm setting
Symbol " " on the display: application is active
4. If necessary: zero the balance/scale
5. Place sample on balance/scale (in this example, 380 g)
6. Store value in memory. Total weight is displayed
steadily;
symbol flashes.
7. Remove sample from balance/scale
8. Place the next sample on the balance/scale
(in this example, 575 g)
9. Store value in memory. Totalized stored weight
is displayed; symbol
Note:
symbol remains on indicating stored
value in memory until cleared
10. To view the current weight for 2 seconds
(if a printer is connected, a printout is generated)
11. Clear totalizing memory
(if a printer is connected, total is printed)
12. End Totalizing
13. Reactivate Totalizing (if no other
application program has been selected)
(if a printer is connected, total is printed).
14. Repeat procedure starting from Step 6.
10
flashes.
Key (or instruction)
(ZERO) > 2 sec
(F) repeatedly
(ENTER)
(ZERO)
(ENTER)
(ENTER)
(F)
(F) > 2 sec
(ENTER) > 2 sec
(F)
Display