Olympus DP12 Instrukcja obsługi - Strona 42
Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Aparat cyfrowy Olympus DP12. Olympus DP12 48 stron. Microscope digital camera system
11
TROUBLESHOOTING GUIDE
Under certain conditions, the performance of the camera system may be adversely affected by factors other than defects. If
problems occur, please review the following list and take remedial action as needed. If you cannot solve the problem after
checking the entire list, please contact your local Olympus representative for assistance.
Trouble
a) The control box does not
work.
b) Picture cannot be recorded
by pressing the EXPOSE but-
ton.
c) Picture cannot be played on
the LCD monitor.
d) The LCD monitor is hard to
view.
e) The picture is not in focus.
Cause
The main switch of the control box is not
ON.
The AC adapter is not connected prop-
erly.
The PC connection cable is connected
to the control box.
Data is being written in the SmartMedia.
The SmartMedia has a write protect seal
attached to it or SmartMedia is not in-
serted.
The SmartMedia is full.
The main switch is not ON or the cam-
era system is not in PLAY mode.
No data has been recorded in the
SmartMedia in use.
The monitor screen is subject to direct
light.
The monitor brightness is not adjusted
properly.
The microscope is not focused properly.
The aperture iris diaphragm of the con-
denser is open too wide.
The field iris diaphragm is not set prop-
erly.
Lens components of the microscope are
contaminated or the cover glass on the
bottom of the camera unit is stained.
Remedy
Set the main switch to " I " (ON).
Connect it properly to the control box and
to the power outlet.
Set the main switch to " " (OFF), unplug
the PC connection cable and retry op-
eration.
Operate the camera system from the util-
ity software running on the PC.
Release the EXPOSE button and wait
until the CARD access indicator stops
blinking in green before pressing the
EXPOSE button again.
Use another SmartMedia.
Use another SmartMedia, erase un-
wanted pictures or download existing
data to your PC then erase all pictures in
the SmartMedia.
Set the main switch to " I " (ON).
Select the PLAY mode.
Check the number of remaining pictures.
Turn off the direct light or screen it with
your hand, etc.
Adjust the brightness.
Adjust the focus correctly with the fine
adjustment knob.
Close the aperture iris diaphragm a little.
Adjust the field iris diaphragm until the
image circumscribes the field of view.
Clean the objective, photography lens,
condenser and/or window lens of the
microscope, or clean the cover glass on
the bottom of the camera unit.
DP12
Page
13
9
33
11
11
13
14
13
–
28
–
–
–
2
39