CMOS color camera SC180
3
System environment
3.1
General remarks
To acquire images with best possible quality, the microscope has to be configured
correctly. The illumination for example must be set in a way to enable image
acquisitions with appropriate exposure times.
This camera installation manual does not contain explanations on how to setup the
microscope. Please refer to the microscope manual for this information. Additionally,
the online help of OLYMPUS Stream and cellSens contains detailed information on
image acquisition.
Using the camera with non-Olympus microscopes
The SC180 can be used in combination with many available microscopes.
However, its performance in combination with non-Olympus microscopes has not
been tested.
Using the SC180 in special application contexts
Application context
Using objectives with
low magnification (<
10x)
Using objectives with
high magnification (>
40x)
Darkfield and fluores-
cence microscopy
10
Explanation
Depending on the microscope configuration,
vignetting might occur in the image if objectives
with a magnification of less than 10x are used.
The maximum exposure time might not be suffi-
cient to acquire a correctly exposed image.
The SC180 is not suitable for darkfield or
fluorescence microscopy due to its limited
exposure range.
Countermeasure
Switch on the shading
correction or use a camera
adapter with a higher magnifi-
cation.
Increase illumination intensity
or use a camera adapter with
a lower magnification.
No countermeasure
available.