4. Condenser Centration
1)
B r ~ n g the
objective
10X
into the light path.
*
I f a specimen i s placed on the circular rotatable stage
without
a
mechanical stage it is recommended t o hold
the peripheries of the specimen with the stage clips
provided to the circular stage.
Swing in the condenser top lens, and b r ~ n g the specimen
into focus.
Stop down the field iris diaphrag~n w ~ t l i knurled ring
a.
A
slightly blurred image of the field diaphragm can now
be seen in the eyepiece. (Fig.
3 )
Flg.
3
4) Move the condenser up and down to locus on the image of the field diaphragm.
*
I f the specimen slide is too thick, i t is sometimes impossible to obtain a sharply-focused
image.
5)
While widening the diameter of the field progressively,
i i s e
the condenser centering
screws
(3)
to bring the diaphragm iniage Into the center of view. (Fig. 3 )
6) Push the analyzer
@
into the light path, and make sure
that both polarizer and analyzer are set at position
"0"
to attain the "Crossed filter" position. Then loosen the
clamping screw
(2)
o f the condenser. (Fig, 4)
7 )
Remove the speclmen o ~ ~ t
of the light path so that a
transparent area comes into the light path. Keeping the
polarizer at the "0" position, rotate the polarirer rota-
.
.
,
.
,
I + + -
& :
,--
j , i
tion ring @ until the optimum extinction is obtained,
/--:,
then clamp the ring. (Fig.
4 )
Fig.
4