Olympus BH2-5RE Kompletny demontaż, czyszczenie i ponowny montaż - Strona 15

Przeglądaj online lub pobierz pdf Kompletny demontaż, czyszczenie i ponowny montaż dla Mikroskop Olympus BH2-5RE. Olympus BH2-5RE 18 stron. Modular revolving nosepiece

Olympus BH2-5RE Kompletny demontaż, czyszczenie i ponowny montaż

Reinstall the Knurled-Rubber Grip Ring

The final assembly step for the turret assembly is to
reinstall the black knurled-rubber grip ring around the
outer perimeter of the revolving turret and secure it in
place with a suitable color-matching adhesive.
Black E6000 is a good choice of adhesive here, since it
will bond well with the metal of the revolving turret as
well as to the black knurled-rubber grip ring.
Additionally, any of the inevitable squeeze-out that is
not completely removed will not be visible. Apply two
dabs of the black adhesive (item S1 of
the outer perimeter of the revolving turret, spaced
approximately 180° apart (see
Figure 42 – Apply black adhesive to revolving turret
Carefully reinstall the knurled-rubber grip ring onto the
outer perimeter of the revolving turret, making sure to
not stretch or damage the grip ring in the process (see
Figure
43).
Figure 43 – Reinstall the knurled-rubber grip ring
Complete Teardown, Cleaning, and Reassembly of the Olympus BH2-5RE Modular Revolving Nosepiece
Appendix
1) onto
Figure
42).
Use dry cotton swabs lightly dampened with isopropyl
alcohol to thoroughly remove any adhesive squeeze-
out (see
Figure
44).
Figure 44 – Remove any adhesive squeeze-out

Ready for Service

The newly reconditioned BH2-5RE modular revolving
nosepiece assembly is now ready to provide many more
years of service (see
Figure
Figure 45 – BH2-5RE ready to be put back into service

Requirements for Periodic Maintenance

Periodic cleaning and application of fresh grease to the
mechanical detent notches in the stationary base is
necessary to minimize wear of the mechanical detents,
thereby maximizing the useful service life of the
BH2-5RE. This can be easily accomplished by simply
removing the protective cover, cleaning and re-greasing
the detent notches, and then reinstalling the protective
cover per the procedures detailed in this document.
The dovetail slide does not need to be removed to
perform this periodic maintenance. If the equipment
sees heavy usage, this service should be performed on a
3
six-month interval
.
3
Field experience has shown that even heavily used scopes (such as those
used in hospitals and clinical lab settings), when lubricated on a six-month
interval, can be expected to provide many years of trouble-free service.
45).
Page 15 of 18
Revision 3