Olympus BH2-CH Substage Montaż - Strona 8
Przeglądaj online lub pobierz pdf Montaż dla Mikroskop Olympus BH2-CH Substage. Olympus BH2-CH Substage 26 stron.
Figure 4 – Remove the condenser slide assembly
Remove the Locking and Centering Thumbscrews
Loosen and remove the condenser-lock thumbscrew
from the centering yoke (see
Figure 5 – Remove the condenser-lock thumbscrew
Loosen and remove the two centering thumbscrews
from the support yoke (see
Figure 6 – Remove the centering thumbscrews
Remove the Slip Plates
The condenser-slide assembly is shown in
Complete Teardown, Cleaning/Lubrication, and Reassembly of the Olympus BH2-CH Substage Assembly
Figure
5).
Figure
6).
Figure
7.
Figure 7 – The condenser-slide assembly (bottom view)
Use a suitable JIS screwdriver to remove the six M2x4
JIS screws securing the three slip plates to the centering
yoke (see
Figure
8).
Figure 8 – Remove screws securing the slip plates
Remove the three loose slip plates (see
careful to not lose any small brass shim washers that
may be beneath the slip plates.
Figure 9 – Remove the three slip plates
Figure
9), being
Revision 1
Page 8 of 26