2. Assembling
the Halogen Lamp Housing
I
Fig. 20
\
Fig. 21
Fig. 22
Fig. 23
Mounting
the Halogen Bulb
(Figs. 20, 21)
I. Remove the socket clamping screw Q) using the Allen screwdriver
provided with the microscope frame. (Fig. 20)
2. Remove the socket from the lamp housing as indicated by the arrow.
3. Holding the bulb with gloves or a piece of gauze, depress the bulb
clamping levers 0, and insert the bulb pins fully into the pinholes 0.
Gently release the bulb clamping levers @to their original positions to
secure the bulb. Whenever you replace the bulb, first turn off the main
switch and wait for bulb, lamp socket and lamp house to cool down.
(Fig. 21)
* Use only the 12V IOOW halogen lamp (PHILIPS #7724 or #7023).
+ Do not touch the halogen
bulb with bare hands. If the bulb
surface is contaminated
with fingerprints
or dust, wipe the bulb
with a soft cloth or gauze.
Mounting
the Collector
Lens and Socket
(Figs. 22,23)
1. Align the collector lens positioning groove 0 with the pin inside the
lamp mount. Then, pull up the collector lens focusing knob 0, and
insert the collector lens until it comes to the end. (Fig. 22)
2. Securely fasten the collector lens with the clamping screw 0.
3. Remount the lamp socket in the reverse order of removal. (Fig. 23)
4. Tighten the socket clamping screw ((iJ in Fig. 20) using the Allen
screwdriver.