Installing the scanning stage with ultrasonic (When using IX83, IX73 or BX51WI)
1
Downloaded from
ManualsNet.com
CAUTION
Fig. 13
3
2
Fig. 14
5
4
Fig. 15
7
6
Fig. 16
search engine
} This section describes the procedure to install the scanning stage with
ultrasonic IX83 as an example. The same instlation procedures apply to
IX73 and BX51WI as well.
As the objective may interfare with the stage depending on the
focus position, be sure to operate carefully.
1. Move the stage to the front side, and remove the screw 1 attached to
the lower level of the stage by using the minus screwdriver.
2. Place the shim rings 3 provided with BX3-SSU over the screw holes 2
on the rear side of the microscope.
3. Place the stage on the microscope so that the screw holes 4 of BX3-SSU
overlaps the shim rings 3. And tighten the holes with the fixising screw
5 by using the Allen wrench provided with the microscope.
4. Move the stage toward the rear side, and tighten the screw holes 6 with
the fixing screws 7.
The fixing screws are provided with BX3-SSU.
BX3-SSU
(Fig. 13-19)
14