BX40
Fig. 19
Fig. 20
Fig. 21
Fig. 22
a
SpeMnefl
Placement
(Figs. 19-21)
+ For Single slide observations,
the maximum
slide dimensions
are 26 x 76 mm, with a thickness
of 0.9-1.4 mm and a cover
glass thickness of 0.17 mm.
+ When observing
very large specimens,
remove the slide holder
and place the specimen
directly on the Stage.
1 Slide Holder for 2 Soecimens
(Fis. 19) j
1. Turn the coarse adjustment knob @ to lower the Stage.
2. Open the spring-loaded curved finger @ on the slide holder and place
one or two specimen slides on the Stage from the front.
3. After placing the slides as far as they will go, gently release the curved
finger.
Slide Holder for Single Specimen
(Figs. 20, 21)
5-l
=
The specimen tan easily be placed by sliding it into the slide holder
6
from the front. (Fig. 20)
m
Using an Oil Immersion
Objective
Adsorption of immersion oil tan Cause the specimen to drift. In such
cases, it is recommended
to use the optional
specimen
Clip @
(BH2-SCB-3) for oil immersion objectives 0. (Fig. 21)
Q
Adjusting
the Tension
of the X-axis
and V-axis Knobs
(Fig. 221
The tension of the X-axis and Y-axis knobs tan be individually adjusted.
Turning the X adjustment ring @ or the Y adjustment ring @ counter-
clockwise
increases tension, and turning them clockwise
reduces
tension. When adjusting the tension, hold the X-axis and Y-axis knobs
to keep them from turning along with the tension adjustment
rings.
+ If the tension
is adjusted
to tight,
a creaking
Sound may be
heard during Stage travel, and the Stage may backtrack.
0