Using Eyepiece Micrometer Disks
Eyepiece micrometer disks can be inserted into WHN10X-H (or WHN10X)
eyepieces.
When the eyepiece in use does not have a diopter adjustment function,
the micrometer disk cannot be brought into focus if you have poor eye-
sight. Should that be the case, adjust the focus with eyeglasses on.
Use 24 mm diam. x 1.5 mm micrometer disks.
Following Fig. 25, turn the built-in micrometer mounting frame ² counter-
clockwise to remove it from the eyepiece and place a micrometer disk @
into the mounting frame. The engraving on the micrometer disk should
face downward into the micrometer mounting frame.
Screw the micrometer mounting frame back into the eyepiece sleeve.
Selecting the Light Path
5
(with the U-TR30-2 or U-SWTR-3)
Slide the light path selector knob @ to select the desired light path.
Light Path
Selector Knob
Pushed in
Middle position
Pulled out
Symbol
Intensity Ratio
100% for binocu-
lar eyepieces
20% for binocular
eyepieces, 80%
for TV/photogra-
phy
100% for TV/
photography
BX41
(Fig. 25)
(Fig. 26)
Applications
Observation of
dark specimens
Observation of
bright specimens,
photography, TV
observation
Photography, TV
observation