Olympus BX46 Instrukcja obsługi - Strona 31

Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Mikroskop Olympus BX46. Olympus BX46 44 stron. Clinical microscope

Olympus BX46 Instrukcja obsługi
Problem
Field of view of one eye does not
match that of the other.
Image shifts when you touch
stage.
X- and Y-axis knobs are too tight
or too loose.
Stroke has reduced.
Lamp
Cause
Common
Interpupillary distance is incorrect.
Common
Diopter adjustment is incorrect.
Common
Different eyepieces are used on
left and right.
Common
Your view is not accustomed to
microscope observation.
Common
Stage is not properly mounted.
Common
Tension of X- and Y-axis knobs is
too high or too low.
Common
Stage guide is deviated.
Remedy
Adjust interpupillary distance.
Adjust diopter.
Change one eyepiece to match
the other so that both sides are
the same type.
Upon looking into eyepieces,
try looking at overall field before
concentrating on specimen range.
You may also find it helpful to
look up and into distance for a
moment before looking back into
microscope.
Clamp stage.
Adjust tension.
Correct deviation as described.
BX46
Page
11
19
33
18
18
18
27