Olympus BX51-P Instrukcja obsługi - Strona 30
Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Mikroskop Olympus BX51-P. Olympus BX51-P 36 stron. Polarizing microscope
Również dla Olympus BX51-P: Informacje (6 strony), Instrukcja obsługi (36 strony)
SPECIFICATIONS
Item
1. Polarizing intermediate
attachment
(U-CPA and U-OPA)
2. Analyzer
(U-AN360P-2)
3. Revolving nosepiece
(U-P4RE)
4. Stage (U-SRP)
27
Polarizing intermediate
attachment (U-CPA)
Field No.
Bertrand lens
Bertrand aperture stop
Bertrand lens knob
position for switch
between orthoscopic
and conoscopic
observation
Analyzer slot
360° dial rotation
Minimum scale reading: 0.1° (Vernier scale)
Type: Centerable quadruple
Attachable compensators: Quarter wave plate (U-TP137), sensitive tint plate (U-TP530)
and all kinds of compensators by means of test plate
adapter (U-TAD).
Type: Polarizing rotatable stage with 3-point centering mechanism 360° horizontal
rotation, clamps at any desired position
360° scale (minimum division:
1°; minimum reading 6' by means of vernier scale)
45° click stop knob
· Stage clips (U-SCB2) attachable for holding specimen in place
· Attachable mechanical stage (U-FMP)
· Universal stage (ZEISS Co. manufacture) attachable
Specifications
Intermediate attachment
22
Focusable
Fixed diaphragm
Pushed in
position: { IN
Pulled out
position: ¦ OUT
Slot for rotatable analyzer (U-AN360P-2)
for AN360P (U-OPA)