Using the ECO/NORMAL Switch
} The ECO mode helps prevent leaving the halogen bulb ON while the
microscope is not used.
1. Press the ECO/NORMAL switch 3 to ON (so that the ECO indicator |
lights in green). Hereafter, when the human sensor 5 of the microscope
has not detected human motion, etc. for more than about 30 minutes,
the lighting halogen bulb is turned OFF automatically (the ECO indicator
| lights in red).
2. To turn the bulb ON again after it has been turned OFF automatically:
· Set the main switch to "
· Press the ECO/NORMAL switch 3 twice.
} In the ECO mode, the human sensor may sometimes be unable to detect
a human body and the halogen bulb be turned OFF during use of the
microscope in the following cases.
· When the heat source (human skin, etc.) in the detection range of
the human sensor hardly moves or moves at a high speed.
· When far-IR rays from the sun or incandescent lamp is directly incident
on the human sensor.
· When an object that transmits little far-IR rays, such as glass or acrylic
plate, is located between the human sensor and the person operating
the microscope.
· When there is little difference between the ambient temperature and
the body temperature of the person.
} In the ECO mode, the halogen bulb may not be turned OFF automatically
in the following cases. In addition to these cases, the halogen bulb may
neither be turned OFF in the case of sensor detection error.
· When the sensor detects the movement of a heat source other than
human body.
· When the sensor detects a heat source that repeats heat generation,
such as a motor repeating ON-OFF.
· When the temperature in the detection area changes suddenly, for
example due to hot/cold winds of an air conditioner or vapor from
a room humidifier.
4
Using the Filters
} You can place a filter in the light path with one of the following methods.
· Push in the knob of the desired built-in filter to engage it in the light
path. (Page 15)
· Place a filter on the filter mount on the base and engage in the light
path. (Page 15)
· Insert a filter in the U-FC filter cassette, attach it on the filter mount
and engage it in the light path by sliding the filter lever. (Page 15)