Olympus BX61 Instrukcja obsługi - Strona 14
Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Mikroskop Olympus BX61. Olympus BX61 40 stron. Motorized system microscopes
‡
‰
10
†
Š
Fig. 9
Fig. 10
‹
Fig. 11
Mounting the Filter Cassette
1. Fully loosen the filter cassette clamping screw †. (Fig. 9)
2. Align the key ‡ on the bottom surface of the filter cassette with the
positioning slot Š on the filter mount, then snap the filter cassette into
place from above.
3. Rotate the filter cassette to align its sides with the base. (Fig. 10)
4. Align the clamping screw † with the positioning hole ‰ on the filter
mount, then tighten the screw to fasten the filter cassette.
# When the filter cassette is installed, the condenser may hit the filter
cassette when the stage is lowered. Therefore, exercise caution
when lowering the stage with the filter cassette installed.
Using the Filter Cassette (Fig. 11)
Usable Filters
45ND-6, 45ND-25
45G-530, 45G-533, 45IF550
45Y-48
45O-560
Up to three of the above filters can be loaded in the filter cassette. Mov-
ing the levers ‹ on the left and right sides of the cassette to the IN
position brings the corresponding filter into the light path.
Applications
Neutral density filter
Green
Yellow
B&W contrast filter
Orange