Olympus BXC-CBB Instrukcja obsługi - Strona 17

Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Mikroskop Olympus BXC-CBB. Olympus BXC-CBB 41 stron. Modular microscope assemblies hardware

1.
Loosen the nosepiece clamping screw (A) of BXC-RLI/BXC-RLI-CA/BXC-RLI-LGCA using the Allen screwdriver provided
with BXC-RLI/BXC-RLI-CA/BXC-RLI-LGCA.
The following steps are described using only BXC-RLI.
NOTE:
Be careful, if the clamping screw is loosened too much, it may come off.
2.
Insert the nosepiece from the front side by aligning the slide dovetail (B) of the nosepiece with the nosepiece mount
dovetail (C) of BXC-RLI, and push it until it touches the end.
NOTE:
Tightening the clamping screw before the nosepiece touches the end may damage the product or cause poor
connection of the connector.
3.
While pushing the nosepiece in the attaching direction with your left hand, hold the Allen screwdriver between the
pointing finger and the thumb and tighten the nosepiece clamping screw (A) to secure the nosepiece.
3.9 Storing the motorized nosepiece connector of the motorized light guide reflected light
illuminator
You can store the motorized nosepiece connector inside BXC-RLI/BXC-RLI-CA/BXC-RLI-LGCA if you are not using it, such
as when using a manual nosepiece.
The following steps are described using only BXC-RLI.
NOTE:
Once you remove the motorized nosepiece connector, its performance cannot be guaranteed even if you restore it.
13
3. Setup procedures
3.9 Storing the motorized nosepiece connector of the