Olympus CHT Instrukcja obsługi - Strona 13
Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Mikroskop Olympus CHT. Olympus CHT 24 stron. Biological microscopes
Również dla Olympus CHT: Instrukcja obsługi (23 strony)
• Cover glass
Use cover glasses of 0.17 mm thickness in conjunction with the objectives marked
wi
th the inscription
"160/0 .17" for op(lmum performance of these objeclfves .
• Specimen slide
Specimen slides between 0.9 mm and 1.2 rnm in thickness are recommended for the CHS/CHT microscopes.
If the thickness of a slide exceeds this range, illumination may sometimes be impaired.
D
Focus
1) S",>,ing in the
lOX
objective.
2) Bring the specimen into focus by means of the coarse and fine
adjustment knobs.
*
Rotate the focus adjustment knobs clockwise (in the direction of
the arrow in Fig. 16), and you can raiS1! the stage (or the specimen
approaches the objective
I.
•
o
OJ
Ul
m
::0
<
»
-l
o
Z
Fig. 16
III
Interpupillarv distance adjustment
1)
Looking through the binocular tube, move the knurled dovetail slides
CU
in the directions of the arrows until a perfect binocular vision
is obtained. (Fig. 17)
2) If you rr,emorize your interpupillary distance setting on the scale
(2) provided between the dovetail slides
CU,
it is convenient
\0
ob-
tain a proper selling next
tin~e.
(Fig. 17)
Fig. 17
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - m J