CX21 MAINTENANCE MANUAL
1. Optical Adjustment
PREPARATION Adjusting the left/right optical axis
*Eyepiece
for CX21
*It is necessary to set the
cross micrometer disk to
the above eyepiece.
Objective 10X
Specimen
whose center is
identified
(concentric circles etc.)
Before adjustment
Cross center of
eyepiece
CHAPTER 3
REPAIR PROCEDURE
If the left/right optical axis is remarkably displaced at
checking, perform the following adjustment.
*Insert the cross eyepiece
Align the specimen center with the cross center of the
eyepiece by turning the control knob of the stage.
Specimen center
into the right sleeve.
After adjustment
- 11 -
REPAIR PROCEDURE
Adjust the interpupillary
distance to about 62mm
(See the illustration)
75
70
60
50
Position of
approx. 62mm
Control knob