Olympus CX33 Instrukcja obsługi - Strona 28
Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Mikroskop Olympus CX33. Olympus CX33 40 stron. Biological microscope
Problem
9.
The stage descends by its own
weight.
Or the focus is lost due to
slippage of the coarse focusing
knob.
10. The field of view is not matched
between two eyes.
11. When changing the objective
from lower magnification to
higher magnification, the
objective collides with the
specimen.
12. The LED illumination does not
light.
13. When changing the
magnification, the specimen is
defocused significantly.
Repair request
If you cannot solve the problems even though taking actions described in "Troubleshooting", please contact Olympus
for assistance. Please provide us the following information at that time.
24
The tension adjustment ring is
loosened too much.
The interpupillary distance is not
adjusted correctly.
The difference in diopter of two eyes
is not corrected properly.
Different eyepieces are used for right
side and left side.
The specimen is placed upside
down.
The cover glass is too thick.
The AC adapter or power cord is
disconnected.
The diopter is incorrect.
Product name and abbreviation (Ex.: Biological Microscope CX33RTFS2)
Product number
Phenomena
Cause
Tighten the tension adjustment ring to
get proper tension.
Adjust it correctly.
Correct it properly.
Use the same eyepiece for right side
and left side.
Place the specimen with the cover
glass side facing up.
Use a cover glass with thickness of
0.17 mm.
Connect the power cord to a power
outlet securely.
Adjust the diopter of eyepieces
correctly.
Remedy
Page
14
15
16
29
10
10
32
16