Aperture Iris Switching
Functions
1. Sets the aperture iris diaphragm to the specified position.
2. Checks the current position of the aperture iris diaphragm.
Special notes
1. The aperture iris diaphragm positions are referred to as; 0 = minimum; 1550 = intermediate; 3113 = maximum.
2. When a position out of the specified range (0 - 3113) is specified, an illegal parameter error is identified and no operation
is executed.
3. The aperture iris diaphragm controlled with these commands is the reflected light aperture iris diaphragm.
Command descriptions
Command
1EASB
n
(n: 0 to 3113)
1EAS?
"
" represents a space.
*
)
Line feed command <CR><LF> is required after commands and line feed command <CR><LF> are added after any
response.
33
Type
Description
Request
Setting of the aper-
ture iris diaphragm
to the position with
the number speci-
fied with "n".
Query
Check of the cur-
rent position of the
aperture iris dia-
phragm.
Response
Result
1EAS
+
Successful
1EAS
!,E01130
Error
1EAS
!,E01120
Error
1EAS
n
Successful
(n: 1 to 3113)
1EAS
!,E01150
Error
1EAS 1EAS?
Description
The aperture iris diaphragm is set
to the specified position.
The connection is disconnected
due to an abnormality with the
aperture iris diaphragm.
A string other than "?" is sent af-
ter "1EAS".
The current position of the aper-
ture iris diaphragm is returned.
A string is sent after "1EAS?
".