2
Adjusting the Diopter
2
(Zoom Confocality Adjustment)
The adjustment operation is variable whether the eyepieces
TIP
use an eyepiece micrometer disks or not.
Eyepieces without eyepiece micrometer disk
1
Turn the diopter adjustment rings b of the left and right eyepieces
to positions "0".
2
Place an easy-to-observe specimen on the stage plate.
3
Set the zoom adjustment knob c for the lowest magnification and
rotate the focus adjustment knob d to bring the specimen into focus.
4
Set the zoom adjustment knob c for the highest magnification and
rotate the focus adjustment knob d to bring the specimen into focus.
5
Set the zoom adjustment knob c for the lowest magnification and
rotate the diopter adjustment rings b of the left and right eyepieces,
instead of the focus adjustment knob, to bring the specimen into focus.
Eyepieces with eyepiece micrometer disk
1
Look through the eyepiece with eyepiece micrometer disk and rotate its
diopter adjustment rings b to bring the micrometer scale into focus.
2
Place an easy-to-observe specimen on the stage plate.
3
Set the zoom adjustment knob c for low magnification, look through
the eyepiece with eyepiece micrometer disk and rotate the focus
adjustment knob d to bring the specimen into focus.
4
Set the zoom adjustment knob c for the highest magnification and
rotate the focus adjustment knob d to bring the specimen into focus.
5
Set the zoom adjustment knob c for the lowest magnification and
rotate the diopter adjustment ring e of the eyepiece without eyepiece
micrometer disk, instead of the focus adjustment knob d , to bring
the specimen into focus.
Note the left and right diopter scale values so that it can be
TIP
quickly duplicated in future observations.
Diopter scale of the 10X eyepieces
The valid range of the diopter scale is between –8 and +5 but
TIP
a slight margin is added to it. Therefore, the diopter value may
exceed +5 or –8 when it is adjusted to the maximum. In this
case, whether the diopter value is over +8 or under –5 can be
identified by the length of the eyepiece.