Olympus SZX16 Instrukcja obsługi - Strona 28
Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Mikroskop Olympus SZX16. Olympus SZX16 40 stron. Research high-class stereo microscope
Również dla Olympus SZX16: Przegląd (24 strony), Instrukcja obsługi (36 strony), Instrukcja obsługi (28 strony)
![Olympus SZX16 Instrukcja obsługi](/uploads/products/server2/4283918/4283918.jpg)
e
d
c
3
Operation
24
Downloaded from
ManualsNet.com
f
a
search engine
4
Place the revolving nosepiece mount c (with its objective mounts d facing
upward) where the objective mount has been by aligning the screw holes.
Using the Allen wrench (for M3 screws), clamp the revolver mount with
the provided three clamping screws (M3, 6 mm long) e .
· Two screw holes are provided at each position with an
NOTE
interval of 5 mm. When not using the SZX2-RFA16 reflected
fluorescent light illuminator, use the screw holes closer to
the front.
· As the screw holes may be hidden behind the objective
mounts d , clamp the nosepiece mount while rotating it.
5
Install the focusing assembly on the pillar, and install the observation
tube in the original position.
6
Attach two objectives onto their respective objective mounts d by
screwing.
Hold the objective and gently rotate it until a click position on the front,
where the objective to be used is engaged in the light path.
Operation for Optical Axis Aligned Photomicrography
Align the center of objective and that of the illumination
TIP
light path with the photo light path (the right line) to enable
photomicrography with high contrast.
1
Rotate the desired objective a clockwise (17°) till the click position
for the photo light path.