4-4 TV Observation and Photomicrography

}When TV observation or photomicrography is required, use the SZX-TR30 trinocular tube.
A TV camera and/or digital camera unit can be mounted on the SZX-TR30 trinocular tube by means of the TV adapter
and/or camera mount adapter*.
* The camera mount adapter is not necessary if a TV adapter equipped with a camera mount is used.
For details, please also read the instruction manuals for the TV adapter and digital camera.
1

Selecting the TV Adapter Magnification

Set the magnification of the TV adapter according to the size of the CCD in the TV camera or digital camera.
The following figures show the TV observation areas when the WHSZ10X eyepieces with FN 22 are used.
When the 0.5X TV adapter is used
²
Fig. 12
1 in. CCD
2/3 in. CCD
1/2 in. CCD
WHSZ10X
FN 22
2
Mounting the TV Adapter
1
1. Using the Allen screwdriver, fully loosen the straight tube clamping screw
@ of the straight tube mount on the top of the trinocular tube.
2. Fit the round dovetail ² of the TV adapter into the straight tube mount of
³
the trinocular tube, and tighten the clamping screw @.
3. Mount the TV camera on the TV adapter. A camera mount adapter may
be required with certain TV adapters.
3
Selecting the TV Camera Light Path
Pull out the light path selector knob ³ to select the Binocular 20%/TV &
Photo 80% light path setting.
When the 1X TV adapter is used
SZX7
(Fig. 12)
(Fig. 12)