Olympus SZX9 Instrukcja obsługi - Strona 28
Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Mikroskop Olympus SZX9. Olympus SZX9 34 stron. Research stereomicroscope system
9-5 BX Stage Adapter Type 1 SZX-STAD1
1
Introduction
This adapter is for installation of a U-SRG or U-SRP rotary stage on the SZX standard base or a SZX series illumination
base. When the U-SRP rotary stage is used together with the U-FMP mechanical stage, X-Y directional movement be-
comes possible, which is convenient for framing during photomicrography.
To cover the height of the stage adapter, we recommend combination with the SZH-P400 auxiliary pillar (together with the
SZX-R drop prevention collar).
2
Mountable Bases, Restrictions
Base
Standard base SZX-ST
· Transmitted illumination base
SZX-ILLK
· Advanced illumination base
SZX-ILLB2
· Brightfield/darkfield transmitted
illumination base
SZX-ILLD2
Usable Objectives
0.5X to 2X
The same restrictions apply regardless of whether the stage adapter is used
or not. (Refer to the SZX illumination base´s instruction manual.)
Note that the field illuminated by the transmitted light is limited depending
on the diameter of the opening of the stage plate in use.
# Darkfield observation is not possible using the SZX-ILLD2.
# The illumination intensity may be reduced when a frosted filter is used.
SZX
Restrictions
None
24